Catalase test

This test detects the catalase enzyme present in most cytochrome-containing aerobic and facultative anaerobic bacteria. The catalase enzyme converts hydrogen peroxide ( H2O2 ) to oxygen and water. This reaction can be observed as the rapid formation of bubbles.

The reaction

2 H2O2 → 2 H2O + O2

Catalase positive organisms

Gram-positive organismsStaphylococcus, Micrococcus,
Aerococcus urinae (Note: Aerococcus viridans is negative),
Listeria, Corynebacterium diphtheriae
Bacillus (to differentiate from aerotolerant Clostridium. which is negative)
Nocardia
Rhodococcus equi.
Mycobacterium tuberculosis
Gram-negative organismsCampylobacter fetus, Campylobacter jejuni, and Campylobacter coli.
Enterobacteriaceae (Citrobacter, E. coli, Enterobacter, Klebsiella, Shigella, Yersinia, Proteus, Salmonella, Serratia)
Pseudomonas
FungusAspergillus, Cryptococcus

Can we get a false-positive or false-negative reaction?

False-positive test

  • Catalase test should not be performed on colonies grown on media containing whole red blood cells (e.g. blood agar). Whole red blood cells may contain catalase; the test could be false positive.
  • Some strains of Aerococcus and Enterococcus may give a false positive reaction due to pseudocatalase formation.
  • Some inoculating loops or wires (nichrome) can react with the hydrogen peroxide to produce false positive reactions.
  • Dirty glass test tubes/ bijoux bottles can also produce false positive results.

False-negative test

  • Older cultures: organisms in older cultures may give a false negative result. The catalase test should ideally be done from culture 18-24 hours old.
  • Anaerobic bacteria – Cultures of anaerobic bacteria should be exposed to air for 30 min before testing
  • Hydrogen peroxide is unstable and must be refrigerated at all times. Avoid any undue exposure to light to prevent a false positive result.
Test methods – multiple methods are available:

Tube or bottle method – put 4-5 drops of H2O2 in a test tube or bijoux bottle. Pick a few colonies with a sterile loop and rub them on the inside wall of the tube above the H2O2 . Cap the bottle/tube. Tilt the bottle to bring H2O2 over the colonies. Look for bubble formation.

Agar slant method – Put 1 ml of H2O2  on a heavily inoculated 18-24 hr old culture on a nutrient agar slant. Replace the cap. Look for bubble formation.

Other methods –
Test on a glass slide or using a capillary tube

Positive result – bubble formation
Negative result – no bubble formation

Safety:
1. The test may produce bacterial-laden aerosol – this test must be performed by a method (tube/bottle or agar slant method) or/and inside a safety cabinet.
2. H2O2 is corrosive; appropriate protection should be taken.
Catalase test (Wikipedia,  Nase)
Disclaimer: This is for academic use only. Managing an infection is complex and depends on taking multiple factors into account. Please speak to your doctor for clinical advice. This article must never be used for clinical purposes.